Available to Buy Online
View Configurations
Configured models contain the following attributes:
Oscilloscope Probes100MHz <= 1GHzPlease use 'Get Quick Quote' or ‘Configure Unit’ and we’ll contact you and quote for your exact need.
You can also get in touch, if you prefer, and our experts will be able to guide you with the right application and procurement solution for this product.
Attribute | Value |
---|---|
High Frequency (range) | 100MHz <= 1GHz |
A high voltage differential probe is used for measuring the voltage difference between two test points where neither test point is at ground. High voltage differential probes from Tektronix can be used for signals up to 6000 V. These probes are the best choice for making non-ground referenced, floating or isolated measurements in large part due to their common mode rejection capability. These products are designed, manufactured, and serviced by Tektronix.
Key Features |
1 GHz and 500 MHz probe bandwidth |
<1 pF differential input capacitance |
1 MΩ differential input resistance |
±42 v (DC + pk AC) differential input voltage |
>18 dB CMRR (at 250 MHz 50X attenuation) |
Benefits | |
Features | Outstanding electrical performance - Selectable bandwidth-limiting filters - DC reject Versatile DUT connectivity - Small compact probe head for probing small geometry circuit elements - Straight pin, square pin, solder down, variable pitch standard accessories - Robust design for reliability Easy to use - Provides automatic units scaling and readout on the oscilloscope display - TDP1000, TDP0500 - Connect directly to oscilloscopes with the TekVPI™ probe interface - Easy access to scope-displayed probe menu for probe setup control and operating status information - Remote GPIB/USB probe control through the oscilloscope - AutoZero – zeros out output offset P6251 - Connect directly to the TekProbe™ interface oscilloscopes, or to TekConnect® oscilloscopes using TCA-BNC adapter |
Applications | High-speed switch mode power supply design CAN/LIN bus design Digital design and characterization Manufacturing engineering test Research and development |
High-Voltage Differential Probes | The TDP1000, TDP0500, and P6251 High-voltage Differential Probes are specifically designed for use with and direct connection to Tektronix oscilloscopes with either the TekVPI™ probe interface, or TekProbe BNC Interface. These probes achieve high-speed signal acquisition and measurement fidelity by solving three traditional measurement challenges: - Outstanding Electrical Performance - Versatile Device-Under-Test Connectivity - Ease of Use |
Specifications | |||
High Voltage Differential Probes | |||
Model Overview | |||
Model | TDP0500 | TDP1000 | P6251 |
Bandwidth (probe only) | 500 MHz | 1 GHz | 1 GHz |
Attenuation | 5X 50X | 5X 50X | 5X 50X |
Differential Mode Input Voltage | ±42 v (DC + pk AC); 30 Vrms | ±42 v (DC + pk AC); 30 Vrms | ±42 v (DC + pk AC); 30 Vrms |
Rise Time (probe only) | <700 ps | ≤350 ps | ≤350 ps |
CMRR | >55 dB at 30 kHz >50 dB at 1 MHz >18 dB at 250 MHz (warranted at 50X attenuation) | >55 dB at 30 kHz >50 dB at 1 MHz >18 dB at 250 MHz (warranted at 50X attenuation) | >55 dB at 30 kHz >50 dB at 1 MHz >18 dB at 250 MHz (warranted at 50X attenuation) |
Maximum Input Voltage (nondestruct) | ±100 V (DC + pk AC) | ±100 V (DC + pk AC) | ±100 V (DC + pk AC) |
Gain Accuracy at DC | ±5% | ±5% | ±5% |
Differential Input Capacitance | ≤1 pF | ≤1 pF | ≤1 pF |
Differential Input Resistance | 1 MΩ | 1 MΩ | 1 MΩ |
Sensitivity/Noise Level | 5X Attenuation: 2 mVrms 10 mVrms (referred to the probe output) 50X Attenuation: 1 mVrms 50 mVrms (referred to the probe output) | 5X Attenuation: 2 mVrms 10 mVrms (referred to the probe output) 50X Attenuation: 1 mVrms 50 mVrms (referred to the probe output) | 5X Attenuation: 2 mVrms 10 mVrms (referred to the probe output) 50X Attenuation: 1 mVrms 50 mVrms (referred to the probe output) |
Propogation Delay | 6.5 ns | 6.5 ns | 6.5 ns |
Common Mode Input Voltage | ±35 v (DC + pk AC); 25 Vrms | ±35 v (DC + pk AC); 25 Vrms | ±35 v (DC + pk AC); 25 Vrms |
Differential Input Offset Range | ±4.25 V (5X attenuation) ±42 V (50X attenuation) | ±42 V (5X or 50X attenuation) | ±42 V (5X or 50X attenuation) |
DC Reject | 5X Attenuation: 0.4 Hz 50X Attenuation: 4 Hz | 5X Attenuation: 0.4 Hz 50X Attenuation: 4 Hz | 5X Attenuation: 0.4 Hz 50X Attenuation: 4 Hz |
Specifications | |
High Voltage Differential Probes | |
Model Overview | |
Model | Description |
TDP1000 | 1 GHz high-voltage differential probe with TekVPI interface. |
TDP0500 | 500 MHz high-voltage differential probe with TekVPI interface. |
P6251 | 1 GHz High-voltage Differential Probe with TekProbe-BNC Level II Interface. |
Specifications | ||||
High Voltage Differential Probes | ||||
Model Overview | ||||
Model | Attenuation | Bandwidth | Input Impedance | Maximum Voltage |
P5200A | 50X / 500X | 50 MHz | 10 MΩ || 2 pF | 500X: ± 1300 V 50X: ± 130 V |
P5202A | 20X / 200X | 100 MHz | 5 MΩ || 2 pF | 200X: ± 640 V 20X: ± 64 V |
P5205A | 50X / 500X | 100 MHz | 10 MΩ || 2 pF | 500X: ± 1300 V 50X: ± 130 V |
THDP0200 | 50X / 500X | 200 MHz | 10 MΩ || 2 pF | 500X: ± 1500 V 50X: ± 150 V |
TMDP0200 | 25X / 250X | 200 MHz | 5 MΩ || 2 pF | 250X: ± 750 V 25X: ± 75 V |
P5210A | 100X / 1000X | 50 MHz | 40 MΩ || 2.5 pF | 1000X: ± 5600 V 100X: ± 560 V |
THDP0100 | 100X / 1000X | 100 MHz | 40 MΩ || 2.5 pF | 1000X: ± 6000 V 100X: ± 600 V |
TDP0500 | 5X / 50X | 500 MHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |
TDP1000 | 5X / 50X | 1 GHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |
P6251 | 5X / 50X | 1 GHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |