The TCP0020, TCP2020, and TCP202A are a family of high-performance, easy-to-use AC/DC current probes designed for use with a variety of oscilloscopes. The TCP0020 is designed for direct connection to oscilloscopes with the TekVPI™ probe interface and the TCP202A is designed for direct connection to oscilloscopes with the TekProbe™ probe interface. The TCP2020 is designed for use with any instrument with BNC inputs and >100 kΩ input termination.
These AC/DC current probes provide sufficient performance to support 50 MHz measurement system bandwidth. The TCP0020 and TCP2020 provide a maximum 20 ARMS measurement range, while the TCP202A provides a maximum 15 ADC + Peak AC measurement range. These probes also provide excellent accuracy to current levels as low as 10 mA, important for meeting today’s challenging current measurement needs.
Key Features:
- Easy-to-Use and Accurate AC/DC Current Measurements
- DC to >50 MHz Bandwidth
- Core Jaw Diameter of 5 mm (0.2 in.)
- Accurately Measures Current Levels as Low as 10 mA per Division
- High Accuracy with Typically Less than 1% DC Gain Error
- Split-core Construction allows Easy Circuit Connection
- Low Noise and DC Drift
- TCP0020
- 20 ARMS Maximum Current Capability
- 100 A Peak Pulse Current Capability
- Connects Directly to Oscilloscopes with the TekVPI™ Probe Interface
- Provides Automatic Units Scaling and Readout on the Oscilloscope Display
- One-button Degauss and AutoZero Control for Ease of Use
- Easy Access to Scope-displayed Probe Menu for Probe Setup Control and Operating Status Information
- Remote Control Capability through TekVPI Oscilloscope
- TCP2020
- 20 ARMS Maximum Current Capability
- 100 A Peak Pulse Current Capability
- 10 mA/mV Sensitivity
- BNC Output Connects to Most Oscilloscopes and Other Measurement Equipment (>100 kΩ Termination Required)
- Degauss Button and Thumbwheel for DC Offset Adjustment
- Powered by an AC Adapter
- TCP202A
- 15 ADC + Peak AC Maximum Current Capability
- 50 A Peak Pulse Current Capability
- Connects Directly to Oscilloscopes with the TekProbe™ Interface
- Provides Automatic Units Scaling and Readout on the Oscilloscope Display
- Degauss Button and Thumbwheel for DC Offset Adjustment
- Safety Certified
Specifications |
|
General |
|
Characteristic | Description |
Bandwidth | DC to ≥50 MHz |
Rise Time | ≤7 ns |
Max Current | 20 ARMS (TCP0020 and TCP2020)
15 ADC + Peak AC (TCP202A)
|
Max Peak Pulse Current | 100 A (TCP0020 and TCP2020)
50 A (TCP202A)
|
Maximum Sensitivity | 10 mA (on oscilloscopes that support 1 mV/div setting) |
Coupling | DC coupling only |
Max Bare-wire Voltage | 150 V CAT II (300 V CAT II Insulated Wire) |