A low voltage single-ended probe is typically used for measuring high-speed, ground referenced signals up to 12 V. These low voltage probes are the best choice for making measurements on high impedance, high frequency circuit elements which require minimal probe loading. Users should select probes with a low input capacitance specification (~1 pF) to minimize the probe’s loading effect on the circuit. A probe with lower input capacitance will offer higher input impedance at higher frequencies.
Selecting the right probe for your application is key to attaining the best signal fidelity in your measurements. Active probes provide truer signal reproduction and fidelity for high-frequency measurements. The small profile and low-mass head make probing dense circuitry by hand fast and easy.
TAP1500 Single-ended Active FET Probe provides excellent high-speed electrical and mechanical performance required for today's digital system designs.
Key Features |
|
Outstanding Electrical Performance | ≥1.5 GHz probe bandwidth
<267 ps rise time
≤1 pF input capacitance
1 MΩ input resistance
-8 V to +8 V input dynamic range
-10 V to +10 VDC input offset range
|
Versatile Mechanical Performance | Small compact probe head for probing small geometry circuit elements
DUT attachment accessories enable connection to SMDs as small as 0.5 mm pitch
Robust design for reliability
|
Easy to Use | Connects directly to oscilloscopes with the TekVPI™ probe interface
Provides automatic units scaling and readout on the oscilloscope display
Easy access to oscilloscope probe menu display for probe status/diagnostic information, and to control probe DC offset
Remote GPIB/USB probe control through the oscilloscope
|
Applications | High-speed Digital Systems Design
Component Design and Characterization
Manufacturing Engineering and Test
Educational Research
Signals with Voltage Swings up to 16 V
|
Specifications |
|
TAP1500
Active Probe 1.5 GHz
|
|
Feature | Parameter |
Attenuation (probe only) | 10X ±2% |
Bandwidth (probe only) | ≥1.5 GHz |
Rise Time (probe only) | <267 ps |
Input Capacitance | <1 pF |
Input Resistance | 1 MΩ |
Linearity | ±8 V (16 Vp-p) |
Input Offset Range | ±10 V |
Maximum Non-Destructive Input Voltage | ±15 V (DC + peak AC) |
Propagation Delay | 5.3 ns |
Probe Head Size | Height: 4.83 mm (0.19 in)
Width: 7.87.1 mm (0.31 in)
Length: 59.18 mm (2.30 in)
|
Cable Length | 1300 mm (51 in) |
Specifications |
|
|
|
|
Low Voltage Single Ended Probes |
|
|
|
|
Model Overview |
|
|
|
|
Model | Attenuation | Bandwidth | Input Impedance | Maximum Voltage |
P6243 | 10X | 1 GHz | 1 MΩ || ≤ 1 pF | ± 8 V |
TAP1500 | 10X | 1.5 GHz | 1 MΩ || ≤ 1 pF | ± 8 V |
P6245 | 10X | 1.5 GHz | 1 MΩ || ≤ 1 pF | ± 8 V |
TAP2500 | 10X | 2.5 GHz | 40 kΩ || ≤ 0.8 pF | ± 4 V |
TAP3500 | 10X | 3.5 GHz | 40 kΩ || ≤ 0.8 pF | ± 4 V |
TAP4000 | 10X | 4 GHz | 40 kΩ || ≤ 0.8 pF | ± 4 V |