Los modelos configurados tienen las siguientes opciones:
Oscilloscopes & Logic AnalyzersOscilloscopes > 3GHz - < 8GHz100MHz <= 1GHz1GHz <= 10GHzUtilice "Presupuesto Rápido" y nos pondremos en contacto con usted y le cotizaremos para su necesidad exacta.
También puede ponerse en contacto, si lo prefiere, y nuestros expertos podrán orientarlo con la aplicación y la solución de adquisición adecuadas para este producto.
Atributo | Valor |
---|---|
High Frequency (range) | 100MHz <= 1GHz |
High Frequency (range) | 1GHz <= 10GHz |
The 6 Series MSO redefines what a Mixed Signal Oscilloscope (MSO) should be. FlexChannel technology enables each channel input to be used as a single analog channel, eight digital logic inputs (with the TLP058 logic probe), or simultaneous analog and spectrum views with independent acquisition controls for each domain. Imagine the flexibility and configurability this provides.
Troubleshoot and validate high-speed designs with bandwidth that starts at 1 GHz and goes up to 10 GHz . Get accurate measurements with low noise and sample rate up to 50 GS/s . See more of your design with 6- and 8-channel models.
With the lowest input noise and up to 8 GHz analog bandwidth, the 6 Series MSO provides the best signal fidelity for analyzing and debugging today's embedded systems with GHz clock and bus speeds. The remarkably innovative pinch-swipe-zoom touchscreen user interface coupled with the industry's largest high definition display and 4 FlexChannel inputs that let you measure one analog or eight digital signals per channel, the 6 Series MSO is ready for today's toughest challenges and tomorrow's too.
Key Features | |
Input Channels | 4 FlexChannel inputs Each FlexChannel provides one analog signal input or eight digital logic inputs with TLP058 logic probe |
Bandwidth | 1 GHz, 2.5 GHz, 4 GHz, 6 GHz, 8 GHz (upgradable) |
Sample Rate (all analog / digital channels) | Real-time: 25 GS/s Interpolated: 2.5 TS/s |
Record Length (all analog / digital channels) | 62.5 Mpoints standard 125, 250, 500 Mpoints, or 1 Gpoints (optional) |
Waveform Capture Rate | >500,000 waveforms/s |
Vertical Resolution | 12-bit ADC Up to 16 bits in High Res mode |
Standard Trigger Types | Edge Pulse Width Runt Timeout Window Logic Setup & Hold Rise/Fall Time Parallel Bus Sequence Visual Trigger Auxiliary Trigger =5 VRMS, 50O, 400 MHz (Edge Trigger only) |
Standard Analysis | Cursors: Waveform, V Bars, H Bars, V&H Bars Measurements: 36 FastFrameTM: Segmented memory acquisition mode with maximum trigger rate >5,000,000 waveforms per second Plots: Time Trend, Histogram and Spectrum Math: basic waveform arithmetic, FFT, and advanced equation editor Search: search on any trigger criteria Jitter: TIE and Phase Noise |
Specifications | |
MSO64 Mixed Signal Oscilloscope | |
Feature | Parameter |
FlexChannel Inputs | 4 |
Maximum Analog Channels | 4 |
Maximum Digital Channels (with optional logic probes) | 32 |
Bandwidth (calculated rise time) | 1 GHz (400 ps) 2.5 GHz (160 ps) 4 GHz (100 ps) 6 GHz (66.67 ps) 8 GHz (50 ps) |
DC Gain Accuracy | 50 Ω: ±2.0% 4 at >2 mV/div (±2.0% at 2 mV/div typical, ±4% at 1`mV/div typical) 50 Ω: ±1.0% 5 of full scale at >2 mV/div, (±1.0% of full scale at 2 mV/div typical, ±2% at 1 mV/div typical) 1 MΩ: ±2.0%4 at >2 mV/div (±2% at 2 mV/div, ±2.5% at 1 mV/div typical and 500`μV/div typical) 1 MΩ: ±1.0%5 of full scale at >2 mV/div, (±1.0% of full scale at 2 mV/div typical, ±1.25% at 1 mV/div and 500`μV/div, typical) |
ADC Resolution | 12 bits |
Vertical Resolution | 8 bits @ 25 GS/s; 8 GHz on all channels 12 bits @ 12.5 GS/s; 4 GHz on all channels 13 bits @ 6.25 GS/s (High Res); 2 GHz on all channels 14 bits @ 3.125 GS/s (High Res); 1 GHz on all channels 15 bits @ 1.25 GS/s (High Res); 500 MHz on all channels 16 bits @ ≤625 MS/s (High Res); 200 MHz on all channels |
Sample Rate | 25 GS/s on all analog / digital channels (40 ps resolution) |
Record Length | 62.5 Mpoints on all analog / digital channels 125 Mpoints on all analog / digital channels optional 250 Mpoints on all analog / digital channels optional 500 Mpoints on all analog / digital channels optional 1 Gpoints on all analog / digital channels optional |
Waveform Capture Rate | >500,000 wfms/s (Peak Detect, Envelope acquisition mode) >30,000 wfms/s (all other acquisition modes) |
Arbitrary/Function Generator (opt.) | 13 predefined waveform types with up to 50 MHz output |
DVM | 4-digit DVM (free with product registration) |
Trigger Frequency Counter | 8-digit frequency counter (free with product registration) |
Specifications | |
MSO64 Mixed Signal Oscilloscope | |
Model Overview | |
Model | Description |
MSO64 | 4 FlexChannels Included Accessories: Four TPP1000 1 GHz probes. Installation and safety manual (translated in English, Japanese, Simplified Chinese ) Embedded Help Front cover with integrated accessory pouch Mouse Power cord Calibration certificate documenting traceability to National Metrology Institute(s) and ISO9001/ISO17025 quality system registration |
Bandwidth Option | |
6-BW-1000 | 1 GHz |
6-BW-2500 | 2.5 GHz |
6-BW-4000 | 4 GHz |
6-BW-6000 | 6 GHz |
6-BW-8000 | 8 GHz |
Note:
For instruments of 4, 6, or 8 GHz bandwidth, consider a BNC-to-SMA adapter to optimize a high bandwidth connection to the oscilloscope.