配置型号包含以下特性:
Oscilloscope Probes100MHz <= 1GHz特性 | 参数 |
---|---|
High Frequency (range) | 100MHz <= 1GHz |
A high voltage differential probe is used for measuring the voltage difference between two test points where neither test point is at ground. High voltage differential probes from Tektronix can be used for signals up to 6000 V. These probes are the best choice for making non-ground referenced, floating or isolated measurements in large part due to their common mode rejection capability.
The THDP0200 is an active differential probe that is capable of safely measuring differential voltages up to ±1500 V. The probe is effective in making measurements in IGBT circuits such as motor drives and power converters.
The THDP0200 supports bandwidths up to 200 MHz and slew rates up to 275 V/ns at 1/250 gain.
Key Features |
Bandwidths up to 200 MHz |
Up to 6000 V differential (DC + pk AC) |
Up to 2300 V common (RMS) |
Overrange Indicator |
Safety Certified |
Switchable Attenuation |
Switchable bandwidth limit |
Benefits | |
Applications | Floating measurements Switching power supply design Motor drive design Electronic ballast design CRT display design Power converter design and service Power device evaluation |
Connectivity | BNC interface (P5200A probes) TekVPI® interface (TMDP and THDP Series probes) TekProbe® interface (P5202A, P5205A, and P5210A Series probes) |
Specifications | |||
High-voltage Differential Probes | |||
Model Overview | |||
Model | TMDP0200 | THDP0200 | THDP0100 |
Attenuation | 25X / 250X | 50X / 500X | 100X / 1000X |
Differential Voltage (Typical) | 250X: ±750 V 25X: ±75 V | 500X: ±1500 V 50X: ±150 V | 1000X: ±6000 V 100X: ±600 V |
Common Mode Voltage (Typical) | ±750 V | ±1500 V | ±6000 V |
Maximum Input Voltage-to-Earth (Typical) | 450 V CAT I 300 V CAT III | 1000 V CAT II 600 V CAT III | 2300 V CAT I 1000 V CAT III |
Bandwidth (Typical) | 200 MHz | 200 MHz | 100 MHz |
Rise Time | <2 ns | <2 ns | <3.6 ns |
Slew Rate (Typical) | <275 V/ns at 1/250 gain | <650 V/ns at 1/500 gain | <2500 V/ns at 1/1000 gain |
Input Impedance at the Probe Tip (Typical) | 5 MΩ || <2 pF | 10 MΩ || <2 pF | 40 MΩ || <2.5 pF |
Typical CMRR (Typical) | DC: > -80 dB 100 kHz: > -60 dB 3.2 MHz: > -30 dB 100 MHz: > -26 dB | DC: > -80 dB 100 kHz: > -60 dB 3.2 MHz: > -30 dB 100 MHz: > -26 dB | DC: > -80 dB 100 kHz: > -60 dB 3.2 MHz: > -30 dB 100 MHz: > -26 dB |
Cable Length (Typical) | 1.5 m | 1.5 m | 1.5 m |
DC Gain Accuracy | ±2% | ±2% | ±2% |
Specifications | ||||
High Voltage Differential Probes | ||||
Model Overview | ||||
Model | Attenuation | Bandwidth | Input Impedance | Maximum Voltage |
P5200A | 50X / 500X | 50 MHz | 10 MΩ || 2 pF | 500X: ± 1300 V 50X: ± 130 V |
P5202A | 20X / 200X | 100 MHz | 5 MΩ || 2 pF | 200X: ± 640 V 20X: ± 64 V |
P5205A | 50X / 500X | 100 MHz | 10 MΩ || 2 pF | 500X: ± 1300 V 50X: ± 130 V |
THDP0200 | 50X / 500X | 200 MHz | 10 MΩ || 2 pF | 500X: ± 1500 V 50X: ± 150 V |
TMDP0200 | 25X / 250X | 200 MHz | 5 MΩ || 2 pF | 250X: ± 750 V 25X: ± 75 V |
P5210A | 100X / 1000X | 50 MHz | 40 MΩ || 2.5 pF | 1000X: ± 5600 V 100X: ± 560 V |
THDP0100 | 100X / 1000X | 100 MHz | 40 MΩ || 2.5 pF | 1000X: ± 6000 V 100X: ± 600 V |
TDP0500 | 5X / 50X | 500 MHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |
TDP1000 | 5X / 50X | 1 GHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |
P6251 | 5X / 50X | 1 GHz | 1 MΩ || ≤ 1 pF | 50X: ± 42 V 5X: ± 4.2 V |