配置型號包含以下特性:
RF & MicrowaveRF Power Sensors20GHz <= 50GHz特性 | 參數 |
---|---|
High Frequency (range) | 20GHz <= 50GHz |
The U8480 series is the fastest USB thermocouple sensor in the world without compromising the best in class linearity of the legacy 8480/N8480 series thermocouple sensors, all in a compact USB based form factor.
Key Features |
Frequency range of 10 MHz to 50 GHz |
Fastest measurement speed for thermocouple sensor, > 900 readings/s (free run/fast buffer mode) |
Best-in-class power linearity of less than 0.8% |
Wide dynamic range from -35 dBm to +20 dBm |
Real time measurement uncertainty (MU) feature |
S-parameter and gamma correction function |
Accurate calibration and traceability to US National Institute of Standards and Technology (NIST) or National Physical Laboratory (NPL) |
BenchVue software license (BV0007B) is now included with your instrument. BenchVue makes it simple to connect, control instruments, and automate test sequences |
Benefits | |
Perform measurements up to 67 GHz | The U8480 series is Keysight’s first power sensor with the capability to measure down to DC (only applicable for U8481A/85A). Now, with an expanded frequency range up to 50 and 67 GHz, the U8480 series allows you to make fast and accurate high frequency power measurements in applications ranging from PNA/MXG calibration to high volume manufacturing. |
Real time measurement uncertainty | The U8480 Series USB thermocouple power sensors now compute this value in real time and at any given point, removing the need for manual MU calculations, shortening test time and providing highly accurate measurements. Users can also display power measurements and their measurement uncertainty simultaneously. |
S-parameter and Gamma correction | Achieve greater accuracy with the S-parameter and gamma correction features, which correct impedance mismatch errors as well as the errors caused by components inserted between the DUT and power sensor. |
Measurements direct from your PC | The U8480 series is supported by the Keysight BenchVue software. BenchVue makes it easy to control your power meter to log data and visualize measurements in a wide array of display options without any programming. Simply connect the sensor to your PC installed BenchVue BV0007B Power Meter/Sensor Control and Analysis app to perform average power measurement and analysis. |
Specifications | |
U8480 Series USB Thermocouple Power Sensors | |
Model Overview | |
Model | Description |
U8481A | USB thermocouple power sensor (Type-N) |
U8485A | USB thermocouple power sensor (3.5 mm) |
U8487A | USB thermocouple power sensor (2.4 mm) |
U8488A | USB thermocouple power sensor (1.85 mm) |
U8489A | USB thermocouple power sensor (1.0 mm) |
Options | |
U8481A-100 | Standard option – 10 MHz to 18 GHz |
U8481A-200 | DC coupled option – DC to 18 GHz |
U8485A-100 | Standard option – 10 MHz to 33 GHz |
U8485A-200 | DC coupled option – DC to 33 GHz |
U8487A-100 | Standard option – 10 MHz to 50 GHz |
U8488A-100 | Standard option – 10 MHz to 67 GHz |
U8488A-200 | DC coupled option – DC to 67 GHz |
U8489A-200 | DC coupled option - DC to 120 GHz |
U8489A-501 | 1.0 mm (f) to N-Type (m) adapter |
Included Accessories | Power sensor cable 5 ft (1.5 m), default cable length Trigger cable, BNC male to SMB female, 50 Ω, 1.5 m Certificate of calibration U8485A: 3.5 mm (f) to N-Type (m) adapter¹ U8487A/88A: 2.4 mm (f) to N-Type (m) adapter¹ |
Note:
¹ Not applicable to U8481A