配置型號包含以下特性:
RF & MicrowaveRF Power Sensors10GHz <= 20GHz特性 | 參數 |
---|---|
High Frequency (range) | 10GHz <= 20GHz |
Keysight U2020 X-series USB power sensors provides the peak and average power measurement capability of a power meter in a compact and portable form factor.
Key Features |
Frequency range of 50 MHz to 18 GHz |
Wide dynamic range of -30 to +20 dBm (peak/gated), -45 dBm to 20 dBm (average only mode) |
30 MHz video bandwidth and single shot real-time capture at 80 Msample/sec per second |
Internal zeroing & calibration |
High measurement speed of up to 25,000 readings/second (free run/fast buffer mode) |
A BenchVue software license (BV0007B) is now included with your instrument. BenchVue makes it simple to connect, control instruments, and automate test sequences |
Benefits | |
Perform measurements direct from your PC | The U2020 X-Series is supported by the Keysight BenchVue software. BenchVue makes it easy to control your power meter to log data and visualize measurements in a wide array of display options without any programming. Simply connect the sensor to your PC installed BenchVue BV0007B Power Meter/Sensor Control and Analysis app to perform complex pulse analysis and standard average power measurement. |
Built in external trigger in and out | External trigger will enable accurate trigger of small signals close to the signal noise floor. With the built-in trigger in/out connection on the Keysight U2020 X-series USB power sensors; you can connect an external trigger signal from a signal source or the device under test directly to the USB sensor through a standard BNC to SMB cable. |
Specifications | ||
U2020 X-Series | ||
Model Overview | ||
Model | U2021XA | U2022XA |
Frequency Range | 50 MHz to 18 GHz | 50 MHz to 40 GHz 50 MHz to 50 GHz (with Option H50) |
Power Range (Normal Mode) | –30 dBm to 20 dBm (50 MHz to <500 MHz) –35 dBm to 20 dBm (500 MHz to 40 GHz) –35 dBm to 8 dBm (> 40 GHz to 50 GHz) | –30 dBm to 20 dBm (50 MHz to <500 MHz) –35 dBm to 20 dBm (500 MHz to 40 GHz) –35 dBm to 8 dBm (> 40 GHz to 50 GHz) |
Power Range (Average Only Mode (1)(2)) | –45 dBm to 20 dBm (50 MHz to 40 GHz) –45 dBm to 8 dBm (> 40 GHz to 50 GHz) | –45 dBm to 20 dBm (50 MHz to 40 GHz) –45 dBm to 8 dBm (> 40 GHz to 50 GHz) |
Damage Level | 23 dBm (average power) 30 dBm (< 1 μs duration) (peak power) | 23 dBm (average power) 30 dBm (< 1 μs duration) (peak power) |
Rise/Fall Time | ≤ 13 ns (3) | ≤ 13 ns (3) |
Maximum Sampling Rate | 80 Msamples/sec, continuous sampling | 80 Msamples/sec, continuous sampling |
Video Bandwidth | ≥ 30 MHz | ≥ 30 MHz |
Single-Shot Bandwidth | ≥ 30 MHz | ≥ 30 MHz |
Minimum Pulse Width | 50 ns (4) | 50 ns (4) |
Basic Accuracy of Average Power Measurement (5) | ≤ ± 0.2 dB or ± 4.5% | ≤ ± 0.3 dB or ± 6.7% |
Maximum Capture Length | 1 s (decimated) 1.2 ms (at full sampling rate) | 1 s (decimated) 1.2 ms (at full sampling rate) |
Maximum Pulse Repetition Rate | 10 MHz (based on 8 samples/period) | 10 MHz (based on 8 samples/period) |
Connector Type | N-type (m) | 2.4 mm (m) |
Specifications | |
U2020 X-Series USB Peak and Average Power Sensors | |
Model Overview | |
Model | Description |
U2021XA | X-Series USB peak and average power sensor, 50 MHz to 18 GHz Included Accessories: Power sensor cable 5 ft (1.5 m), default cable length BNC male to SMB female trigger cable, 50 Ω, 1.5 m (Quantity: 2) Certificate of calibration |
U2022XA | X-Series USB peak and average power sensor, 50 MHz to 40 GHz Included Accessories: Power sensor cable 5 ft (1.5 m), default cable length BNC male to SMB female trigger cable, 50 Ω, 1.5 m (Quantity: 2) Certificate of calibration |
U2022XA-H50 | X-Series USB peak and average power sensor, 50 MHz to 50 GHz Included Accessories: Power sensor cable 5 ft (1.5 m), default cable length BNC male to SMB female trigger cable, 50 Ω, 1.5 m (Quantity: 2) Certificate of calibration |