Kenmerk | Waarde |
---|---|
Form Factor | Benchtop |
The compact, versatile R&S®SMB100A microwave signal generator with a frequency range up to 40 GHz provides outstanding spectral purity and high output power. In addition, it features easy operation, comprehensive functionality and low cost of ownership.
For even higher frequencies, the frequency range of the R&S®SMB100A can be easily extended with the R&S®SMZ frequency multiplier. The R&S®SMB100A plus R&S®SMZ solution combines easy handling with precise, adjustable output levels in the frequency range from 50 GHz to 110 GHz (adjustable output levels are not possible with R&S®SMZ170).
Key Features |
Wide frequency range from 9 kHz to 12 GHz or from 100 kHz to 40 GHz |
Excellent signal characteristics with low SSB phase noise of typ. –128 dBc (at 1 GHz, 20 kHz offset) |
High output power of typ. up to +27 dBm |
All important analog modulations with AM, FM/φM and pulse modulation supported |
Compact size with only two height units and low weight |
Specifications | ||
SMB100A Microwave Signal Generator | ||
Frequency | ||
Frequency range | R&S®SMB-B112/R&S®SMB-B112L | 100 kHz to 12.75 GHz |
R&S®SMB-B120/R&S®SMB-B120L | 100 kHz to 20 GHz | |
R&S®SMB-B131 | 100 kHz to 31.8 GHz | |
R&S®SMB-B140/R&S®SMB-B140L/R&S®SMB-B140N | 100 kHz to 40 GHz | |
Setting time | SCPI mode | < 3 ms |
List mode | < 1 ms | |
Level | ||
Maximum output power | R&S®SMB-B112/-B112L, 1 MHz < f ≤ 12.75 GHz | +18 dBm |
R&S®SMB-B120 with R&S®SMB-B31, 50 MHz < f ≤ 20 GHz | +16 dBm | |
R&S®SMB-B120L with R&S®SMB-B31, 100 MHz < f ≤ 20 GHz | +19 dBm | |
R&S®SMB-B131/-B140/-B140N with R&S®SMB-B32, 50 MHz < f ≤ 40 GHz | +19 dBm | |
R&S®SMB-B140L with R&S®SMB-B32, 50 MHz < f ≤ 40 GHz | +16 dBm | |
Minimum specified output power | R&S®SMB-B112/-B131/-B140/-B140N | -120 dBm |
R&S®SMB-B112L | -5 dBm | |
R&S®SMB-B120L/-B140L | 0 dBm | |
Setting time (without switching of the mechanical attenuator) | SCPI mode | < 2.5 ms |
List mode | < 1 ms | |
Spectral Purity | ||
SSB phase noise | f = 1 GHz, carrier offset = 20 kHz, 1 Hz measurement bandwidth | < –122 dBc, typ. –128 dBc |
f = 10 GHz, carrier offset = 20 kHz, 1 Hz measurement bandwidth | < –102 dBc, typ. –108 dBc | |
Harmonics | ||
R&S®SMB-B112/-B112L | 1 MHz < f ≤ 6 GHz; level ≤ 13 dBm 1) f > 6 GHz; level ≤ 10 dBm 1) | < –30 dBc |
R&S®SMB-B120/-B120L/-B131/-B140/ -B140N/-B140L | standard; level ≤ 8 dBm 1) | |
f > 1 MHz | < –30 dBc | |
with R&S®SMB-B25/-B26 low harmonic option, low harmonic filter on, level ≤ 10 dBm 1) | ||
1 MHz < f ≤ 150 MHz | < –30 dBc | |
150 MHz < f ≤ 3 GHz | < –58 dBc | |
3 GHz < f ≤ 20 GHz | < –50 dBc | |
f > 20 GHz | < –60 dBc (meas.) | |
Supported Modulation Modes | ||
AM | standard | |
AM depth | 0 % to 100 % | |
FM/φM | standard | |
Maximum FM deviation | f = 10 GHz | 32 MHz |
Maximum φM deviation | f = 10 GHz | 320 rad |
Pulse | R&S®SMB-K21/-K22 pulse modulator | |
Rise/fall time | < 20 ns, typ. < 5 ns | |
ON/Off ratio | > 80 dB | |
Minimum pulse width of pulse generator output | 10 ns | |
Connectivity | ||
Remote control | GPIB IEEE-488.2 Ethernet (TCP/IP), USB | |
Peripherals | USB |
Note:
(1) Or maximum specified output power, whichever is lower.
Specifications | ||
SMB100A Microwave Signal Generator | ||
Ordering Information | ||
Model | Order Number | Description |
Base Unit | ||
SMB100A | 1406.6000.02 | RF and Microwave Signal Generator (1) Including power cable, quick start guide and CD-ROM (with operating and service manual) |
Frequency Options | ||
SMB-B101 | 1407.2509.02 | 9 kHz to 1.1 GHz |
SMB-B102 | 1407.2609.02 | 9 kHz to 2.2 GHz |
SMB-B103 | 1407.2709.02 | 9 kHz to 3.2 GHz |
SMB-B106 | 1407.2909.02 | 9 kHz to 6 GHz |
SMB-B112 | 1407.2109.02 | 100 kHz to 12.75 GHz, with electronic step attenuator |
SMB-B112L | 1407.2150.02 | 100 kHz to 12.75 GHz, without step attenuator |
SMB-B120 | 1407.2209.02 | 100 kHz to 20 GHz, with mechanical step attenuator |
SMB-B120L | 1407.2250.02 | 100 kHz to 20 GHz, without step attenuator |
SMB-B131 | 1407.2280.02 | 100 kHz to 31.8 GHz, with mechanical step attenuator |
SMB-B140 | 1407.2309.02 | 100 kHz to 40 GHz, with mechanical step attenuator |
SMB-B140L | 1407.2350.02 | 100 kHz to 40 GHz, without step attenuator |
SMB-B140N | 1407.2380.02 | 100 kHz to 40 GHz, with mechanical step attenuator, minimum pulse width limited |
Note:
(1) The base unit must be ordered together with an R&S®SMB-B101/-B102/-B103/-B106/-B112/-B112L/-B120/-B120L/-B131/-B140/-B140L/-B140N frequency option.