配置型号包含以下特性:
RF & MicrowaveRF Power Sensors请参考资源中的资料文档或查看产品总览。
如果您想了解更多,请 联系,我们的团队将为您提供帮助。
The NRQ6 is based on receiver technology and can perform band-limited power measurements down to -130 dBm. The NRQ6 combines high-precision with high-speed measurements beyond the limits of currently available power meters. In addition to conventional continuous average measurements, the NRQ6 has a trace display function and also performs ACLR measurements. The optional NRQ6-K1 I/Q data interface, NRQ6-K2 power servoing and NRQ6-K3 phase coherent measurements options complete the NRQ6 as new category of test and measurement devices.
Key Features |
Frequency selective power measurements |
Frequency range: 50 MHz to 6 GHz |
Power measurement range: –130 dBm to +20 dBm |
Automatic frequency and bandwidth detection |
100 MHz measurement bandwidth |
Continuous average, trace and ACLR measurements |
I/Q data capturing for RF vector signal analysis |
Fast power servoing with R&S®SGT100A |
Phase coherent measurements for low-power modulated signals |
Benefits | |
Diverse Measurement Functions | Continuous average power measurements down to -130 dBm I/Q data capturing for RF vector signal analysis Phase coherent measurements Trace measurements ACLR measurements Ultrafast triggered pulse measurements RF power, EVM and phase measurements with one single instrument |
Easy Operation | Intuitive web GUI Diverse autoset functions Automatic frequency tracking Spectrum display for signal check Hardware interfaces for user convenience |
I/Q Data Capturing | The NRQ6 can be used as a standalone RF front end to capture vector-modulated I/Q signals. With the optional NRQ6-K1 I/Q data interface, captured I/Q data can be read out using SCPI commands. The data is demodulated and analyzed using external software, e.g. VSE. With the NRQ6 version 2.10 and the VSE version 1.70, direct data aquisition and control is possible. |
Fast Power Servoing with SGT100A | In component test applications non-linear characteristics of DUT's make it difficult to set up a specific output power. The NRQ6-K2 power servoing option enables a dedicated serial communications channel over the trigger connectors between NRQ6 and SGT100A. The desired power will be set in several iterations and is finished typically in 1 ms to 1.5 ms. |
Phase Coherent Measurements | The synchronous phase coherent measurements (option NRQ6-K3) are based on a primary/secondary relationship between two or more NRQ6. The primary forwards the local oscillator signal (LO) and its clock signal (CLK) to the secondary(ies) and triggers and secondary(ies). The RF phases can be measured at up to 8 ports, where the typical uncertainty is only 0.1 degree at an power of -60 dBm |
High Measurement Speed | Every diode based power meter works without band limitation in the specified frequency range. Precise measurement of low-power signals requires averaging of multiple measurement samples and this increases the measurement time. The diode based power meter technology is mature, and therefore only limited product innovations are possible. The receiver technology based NRQ6 has significantly increased the measurement speed for low-power signals. |
Specifications | ||||||
NRQ6 Frequency Selective Power Sensor | ||||||
Model Overview | ||||||
Model | Order Number | Frequency Range | Level Range | Dynamic Range (sensor dependent) | LAN Interface | Connector Type |
NRQ6 | 1421.3509.02 | 50 MHz - 6 GHz | -130 dBm - +20 dBm | up to 150 dBm (RBW dependent) | Yes | N (m) |
Specifications | ||
NRQ6 Frequency Selective Power Sensor | ||
Model Overview | ||
Model | Order Number | Description |
NRQ6 | 1421.3509.02 | Frequency Selective Power Sensor |
NRQ6-K1 | 1421.4705.02 | I/Q Data Interface |
NRQ6-K2 | 1421.4740.02 | Power Servoing |
NRQ6-K3 | 1421.4770.02 | Phase Coherent Measurements |