配置型号包含以下特性:
RF & Microwave Network AnalyzersEMC / EMI / RFI Analyzers20GHz <= 50GHzDC <= 1kHz特性 | 参数 |
---|---|
High Frequency (range) | 20GHz <= 50GHz |
Low Frequency (range) | DC <= 1kHz |
The only device that offers genuine 3-in-1 functionalities: broadband measurements (DC to 60 GHz), frequency spectrum analysis (DC to 400 kHz) and static field measurements.
The SMP2 field meter provides the most advanced features and technology on the market, designed for users who need high-quality measurements, ease of use and good reporting tools.
Key Features |
3-in-1 Field Meter: spectrum analysis (DC – 400 kHz), broadband (DC - 60 GHz) and static field measurements. |
Real-time FFT-based spectrum analysis using digital processing. |
Weighted Peak Method (WPM) for automatic and real time comparison with limits. |
Special auto-gain function for peak detection (indispensable for procedures such as welding). |
Display of results in RMS and peak values. |
Function for time domain measurement of a specific frequency. |
Automatic field probe recognition. |
E&H field selection. |
Total field value and value for each component (X, Y, Z). |
Frequency SPAN selection. |
High Pass Filters at 1 Hz, 10 Hz, 25 Hz and 100 Hz. |
Max-hold function. |
Display of maximum, minimum and average values. |
Sliding window function according to standards. |
Spatial average function. |
Adjustable measurement time from 1s to 100 hours. |
Programmable measurement to start at a given time. |
Graphic display of broadband measurement changes over time. |
Adjustable graph display from 1 s to 12 minutes. |
Programmable alarm function. |
USB and fibre optics connection. |
Large storage capacity. |
Screen-shot function for reporting. |
PC software for control and reporting. |
Results can be uploaded automatically to a web-page control centre. |
Available in several languages. |
Benefits | |
Broadband Measurements DC - 60 GHz | With RMS and isotropic field probes |
Spectrum Analysis DC | Real-time FFT |
Weighted Peak Method (WPM) | Real-time comparison in % with the selected limits |
Assessments According to European Directive 2013/35/EU | In accordance with international standards: ICNIRP, IEC, EN, IEEE, etc. |
High Performance GPS | Integrated in the instrument |
USB and Fiber Optic Communication | PC software included |
Specifications | |
SMP2 Electromagnetic Field Meter | |
Feature | Parameter |
Versions | Broadband: For broadband measurements using the following probes: WPFx, WPT, WP50, WPH60 and WPH1000. Selective: For frequency selective measurements from 0 to 400 kHz using WP400, WP400-3, WP400c and WPH-DC. Dual: For both kinds of measurements using all field probes |
Field Probes | Automatic detection and recognition |
Broadband | 0 Hz – 60 GHz (depending on field probe) |
Spectrum Analysis | up to 400 kHz |
Weighted Peak Method | 1 Hz – 400 kHz (Real time WPM for direct comparison with limits) |
Readout Values | Total field (instantaneous, max., min. and average) Field components (X, Y, Z) |
E Field Units | V/m, kV/m, μW/cm² , mW/cm² , W/m² , % |
H Field Units | nT, μT, mT, T, A/m, %, mG, G |
Log Time | Configurable (from 0.5 s to 6 min) |
Average Model | Fixed o Sliding, according to international standards |
Average Intervals | 10 s 15 s 30 s 1 min 2 min 5 min 6 min 10 min 15 min 30 min |
Schedule Measurement | Customized (up to 24 hours) |
Memory Capacity | More than 1 million samples |
Data Downloading | Mini-USB and Fibre Optics |
Firmware Updating | Mini-USB |
Alarm | 2400 Hz audible signal (adjustable threshold) |
Display Type | Color transmissive TFT (480 x 272 pixels) |
GSP (optional) | Built-in u-blox 7 (56 independent tracking channels) |
Battery | Internal rechargeable Li-ion |
Autonomy | > 24 hours |
Temperature Range | -10 ºC to +50 ºC |
Humidity | 5% to 95%, non-condensing |
Size | 100 x 215 x 40 mm (3.9 x 8.4 x 1.5 “) |
Weight | Broadband: 560 g (19.7 oz.) Selective: 635 g (22.4 oz.) Dual: 635 g (22.4 oz.) |
Note:
Product specifications and descriptions in this document subject to change without notice.
Specifications | |
Frequency Range of Compatible Field Probes | |
Model Overview | |
Model | Frequency Range |
WPH-DC Selective & Broadband | 0 – 40 kHz |
WP400 Selective & Broadband | 1 Hz – 400 kHz |
WP400-3 Selective & Broadband | 1 Hz – 400 kHz |
WP400c Selective & Broadband | 1 Hz – 400 kHz |
WP50 | 10 Hz – 3 kHz |
WPH60 | 300 kHz – 60 MHz |
WPH1000 | 30 MHz – 1000 MHz |
WPF3 | 100 kHz – 3 GHz |
WPF3-HP | 100 kHz – 3 GHz |
WPF6 | 100 kHz – 6 GHz |
WPF6-HP | 100 kHz – 6 GHz |
WPF8 | 100 kHz – 8 GHz |
WPF8-HP | 100 kHz – 8 GHz |
WPF18 | 300 kHz – 18 GHz |
WPF18-HP | 300 kHz – 18 GHz |
WPF40 | 1MHz – 40 GHz |
WPF60 | 1MHz – 60 GHz |
WPF60S | 1MHz – 60 GHz |
WPT | Selective: 700 – 900, 1800 – 1900, 2100, 2600 Hz |
WP-WIFI | WiFi 2.45 GHz |