配置型号包含以下特性:
Oscilloscope Probes1MHz <= 10MHz特性 | 参数 |
---|---|
High Frequency (range) | 1MHz <= 10MHz |
A high voltage single-ended probe is typically used for measuring ground referenced signals up to 40 kV. However, some single-ended probes are designed for oscilloscopes with isolated or floating inputs for measurements that are not ground referenced. Users should select probes with a low input capacitance specification (< 4 pF) to minimize the probe’s loading effect on the circuit because a probe with lower input capacitance will offer higher input impedance at higher frequencies.
Safe high voltage probe solutions for ground-referenced voltage testing. It is critical to safely and accurately capture real-time signal information from "elevated" or "floating" voltage systems. Our portfolio of high voltage probes provides the safety you need for single-ended, differential or isolated measurements.
For heavy-duty high-performance measurements of voltages over 2.5 kV, the P6015A is the industry standard. You can measure DC voltages up to 20 kVRMSand pulses up to 40 kV (peak, 100 ms duration). The 75 MHz bandwidth enables you to capture fast, high-voltage signals.
The P6015A uses an environmentally safe silicone compound for a dielectric and never needs refilling. Other features include: A 7-49 pF compensation range, small compensation box that fits on adjacent amplifier inputs, and a readout option for use with most Tektronix digital scopes. With the readout option, displayed voltage amplitude values will be the actual signal value rather than understated by a factor of 1,000.
Key Features |
DC to 75 MHz |
High Voltage - 20 kV DC / 40 kV Peak (100 ms Pulse Width) |
Optional 1000X readout coding |
Applications |
Power supply design |
Design motor drive |
Electronic ballast |
Power semiconductors |
Switch mode control |
UPS systems |
Power converters |
Specifications | ||||||
Passive High Voltage Probes | ||||||
Model Overview | ||||||
Model | P5100A | TPP0850 | P5122 | P5150 | P6015A | P6015A Opt. 1R |
Nominal Length | 2 m | 1.3 m | 1.2 m | 2 m | 3 m | 3 m |
Attenuation | 100X | 50X | 100X | 50X | 1000X | 1000X |
Bandwidth | 500 MHz | 800 MHz | 200 MHz | 500 MHz | 75 MHz | 75 MHz |
Rise Time (typ) | <700 ps | <525 ps | 2.2 ns | <700 ps | 4.0 ns | 4.0 ns |
Loading | 40 MΩ/ 2.5 pF | 40 MΩ/ 1.8 pF | 100 MΩ/ 4.0 pF | 40 MΩ/ 3.8 pF | 100 MΩ/ 3.0 pF | 100 MΩ/ 3.0 pF |
Maximum Input Voltage (DC or RMS) | 2.5 kVPeak 1000 VRMS CAT II | 2.5 kVPeak 1000 VRMS CAT II | 1000 VRMS CAT II | 2.5 kVPeak 1000 VRMS CAT II | 20 kV | 20 kV |
Maximum Float Voltage | N/A | N/A | 600 VRMS CAT II | 600 VRMS CAT II 300 VRMS CAT III | N/A | N/A |
Compensation Range (pf) | 7 to 30 | N/A | 10 to 22 | 10 to 25 | 7 to 49 | 7 to 49 |
Readout | Yes | Yes | No | No | No | Yes |
EMC (Compliance Information) | UL61010-031 EN61010-031 IEC61010-031 CSA61010-031 | UL61010-031 EN61010-031 IEC61010-031 CSA61010-031 | UL61010-1 UL61010B-2-031 CSA61010-1 | UL61010-031 EN61010-031 IEC61010-031 CSA61010-031 | UL3111-1 EN61010-1 IEC61010-2-031 CSA1010.1 CSA1010.2.031 | UL3111-1 EN61010-1 IEC61010-2-031 CSA1010.1 CSA1010.2.031 |