Cette configuration a les caractéristiques suivantes:
RF & Microwave Spectrum AnalyzersSpectrum Analyzers-Performance20GHz <= 50GHzDC <= 1kHzBenchtopCliquez sur 'Configurez votre produit' et nous vous contacterons pour vous adresser un devis conforme à votre besoin.
Vous pouvez cliquer sur Contactez-moi, si vous préférez, et nos experts vous guideront vers la bonne solution technique et financière.
Attribut | Valeur |
---|---|
High Frequency (range) | 20GHz <= 50GHz |
Low Frequency (range) | DC <= 1kHz |
Form Factor | Benchtop |
The FSV and the FSVA are a family of versatile signal and spectrum analyzers for users working in the development, production, installation and servicing of RF systems.
The FSV is the ideal instrument for all general-purpose measurement tasks – on the bench, in production, and in the field. It provides digital modulation analysis for the latest cellular and wireless standards with up to 160 MHz analysis bandwidth for measurements on components, chipsets and base stations.
Typical measurement tasks include standard-compliant spectrum emission mask measurements as well as spurious emission and adjacent channel leakage ratio (ACLR) measurements.
Measurement applications for EMI diagnostics, phase noise, noise figure, analog demodulation and vector signal analysis complete the range of functions offered by the versatile R&S®FSV.
Key Features |
Frequency range up to 4/7/13.6/30/40 GHz |
Up to 160 MHz signal analysis bandwidth |
Convenient, intuitive operation with touchscreen based user interface |
0.4 dB level measurement uncertainty up to 7 GHz |
–110 dBc (1 Hz) phase noise at 10 kHz frequency offset |
+15 dBm third-order intercept (TOI) |
Displayed average noise level (DANL) in 1 Hz bandwidth: –155 dBm at 1 GHz, –147 dBm at 30 GHz; with R&S®FSV-B24 preamplifier: –162 dBm at 30 GHz |
General-purpose measurement applications for phase noise, noise figure, vector signal analysis, analog demodulation, EMI diagnostics |
Wireless measurement applications for LTE (including LTE-Advanced), WLAN (including IEEE 802.11ac), WCDMA/HSPA+, TD-SCDMA, GSM/EDGE, CDMA2000®/1xEV-DO, Bluetooth® |
Frequency range up to 500 GHz with harmonics mixers |
Keeping test data confidential with removable solid state or hard disk drives |
Specifications | |
FSV Series Signal and Spectrum Analyzer | |
Model Overview | |
Feature | Parameter |
Frequency Range |
DC Coupled: 10 Hz to 4 GHz AC Coupled: 1 MHz to 4 GHz Frequency Resolution: 0.01 Hz DC Coupled: 10 Hz to 7 GHz AC Coupled: 1 MHz to 7 GHz Frequency Resolution: 0.01 Hz DC Coupled: 10 Hz to 13.6 GHz AC Coupled: 10 MHz to 13.6 GHz Frequency Resolution: 0.01 Hz FSV30 DC Coupled: 10 Hz to 30 GHz AC Coupled: 10 MHz to 30 GHz Frequency Resolution: 0.01 Hz DC Coupled: 10 Hz to 40 GHz AC Coupled: 10 MHz to 40 GHz Frequency Resolution: 0.01 Hz |
Interface | GPIB or LAN interface |
Memory Length | max. 200 Msample I and Q |
Impedance | 50 Ω |
Connector | FSV4: N female FSV7: N female FSV13: N female FSV30: APC 3.5mm/N female test port adapter FSV40: 2.92 mm (K)/N female test port adapter |
Display | Display: 21 cm LC TFT color display (8.4") Resolution: 800 × 600 pixel (SVGA resolution) Pixel Failure Rate: < 1 × 10⁻⁵ |
Data Storage | Internal: (Standard) hard disk ≥ 40 Gbyte (nom.) Internal: (with FSV-B18 Option) solid-state drive ≥ 8 Gbyte (nom.) External: supports USB-2.0-compatible memory devices |
Temperature | Operating Temperature: +5 °C to +40 °C Storage Temperature: –40 °C to +70 °C Climatic Loading: +40 °C at 90 % rel. humidity, in line with EN 60068-2-30 |
Power Supply | AC Supply: 100 V to 240 V, 3 A to 1.25 A; 50 Hz to 400 Hz, class of protection I, in line with VDE 411 |
Power Consumption | FSV4: 90 W (nom.), max. 180 W with all options FSV7: 90 W (nom.), max. 180 W with all options FSV13: 115 W (nom.), max. 180 W with all options FSV30: 115 W (nom.), max. 180 W with all options FSV40: 115 W (nom.), max. 180 W with all options |
Safety | in line with EN 61010-1, IEC 61010-1, UL 61010-1, CAN/CSA-C22.2 No. 61010-1 |
Test Mark | VDE, GS, CSA, CSA-NRTL |
Dimensions (WxHxD) | 412 mm × 197 mm × 417 mm 16.22 in × 7.76 in × 16.42 in |
Net Weight (without options (nom)) | FSV4: 9.5 kg (20.94 lb) FSV7: 9.5 kg (20.94 lb) FSV13: 10.3 kg (22.7 lb) FSV30: 10.7 kg (23.58 lb) FSV40: 11.1 kg (24.46 lb) |
Specifications | ||
FSV Series Signal and Spectrum Analyzer | ||
Model Overview | ||
Model | Order Number | Description |
FSV4 | 1321.3008.04 | Signal and spectrum analyzer, 10 Hz to 4 GHz |
FSV7 | 1321.3008.07 | Signal and spectrum analyzer, 10 Hz to 7 GHz |
FSV13 | 1321.3008.13 | Signal and spectrum analyzer, 10 Hz to 13.6 GHz |
FSV30 | 1321.3008.30 | Signal and spectrum analyzer, 10 Hz to 30 GHz |
FSV40 | 1321.3008.40 | Signal and spectrum analyzer, 10 Hz to 40 GHz |
FSV40 * | 1321.3008.39 | Signal and spectrum analyzer, 10 Hz to 40 GHz |
Note:
* Max. bandwidth 10 MHz.
Specifications | ||
FSVA Series Signal and Spectrum Analyzer | ||
Model Overview | ||
Model | Order Number | Description |
FSVA4 | 1321.3008.05 | Signal and spectrum analyzer, 10 Hz to 4 GHz |
FSVA7 | 1321.3008.08 | Signal and spectrum analyzer, 10 Hz to 7 GHz |
FSVA13 | 1321.3008.14 | Signal and spectrum analyzer, 10 Hz to 13.6 GHz |
FSVA30 | 1321.3008.31 | Signal and spectrum analyzer, 10 Hz to 30 GHz |
FSVA40 * | 1321.3008.41 | Signal and spectrum analyzer, 10 Hz to 40 GHz |
Note:
* Max. bandwidth 10 MHz.