The 5500A is a revolutionary product that addresses a wide cross-section of your electrical calibration work load. It sources direct voltage and current, alternating voltage and current with multiple waveforms and harmonics, two simultaneous voltage outputs or voltage and current, and simulates power with phase control, resistance, capacitance, thermocouples and RTDs. The 5500A's Oscilloscope Calibration options provide level sine wave, fast edge, time mark and amplitude signals for calibration of oscilloscopes up to 600 MHz.
Key Features |
Handheld and bench multimeters |
Oscilloscopes and ScopeMeter® Test Tools |
Wattmeters |
Analog volt/ohm/amp/watt instruments |
Electronic thermometers |
Data loggers |
Strip chart recorders |
XY Recorders |
Power harmonics analyzers |
Process calibrators |
Current clamps |
And related instruments |
Specifications |
|
5500A Multi-Product Calibrator |
|
Features | Parameters |
Voltage DC | Range:
0 to ±1020V
Best 1 Year Specification:
±50 ppm of setting
|
Current DC | Range:
0 to ±11A
Best 1 Year Specification:
±0.01% of setting
|
Resistance | Range:
0 to 3.29.999 MΩ
Best 1 Year Specification:
±0.009% of setting
|
Voltage AC | Range:
1 mV to 1020V
10 Hz to 500 kHz, Sine
Best 1 Year Specification:
±0.03% of setting
|
Current AC | Range:
29 m A to 11A
10 Hz to 10 kHz, Sine
Best 1 Year Specification:
±0.06% of setting
|
Capacitance | Range:
0.33 nF to 1.1 mF
Best 1 Year Specification:
±0.25% of setting
|
| Range:
-250°C to +2316°C (11 types)
Best 1 Year Specification:
± 0.14°C of setting
|
RTD Source | Range:
-200°C to +630°C
(8 types)
Best 1 Year Specification:
±0.03°C of setting
|
DC Power | Range:
109 mW to 11 kW
Best 1 Year Specification:
± 0.08% of setting
|
AC Power | Range:
109 mW to 11 kW
45-65 Hz, PF=1
Best 1 Year Specification:
± 0.15% of setting
|
Phase | Range:
0 to ± 179.99°
Best 1 Year Specification:
±0.15°
|
Frequency | Range:
0.01 Hz to 2.0 MHz
Best 1 Year Specification:
±25 ppm of setting
|
Specifications |
|
5500A Multi-Product Calibrator |
|
Features | Parameters |
Warm up Time | Twice the time since last warmed up, to a maximum of 30 minutes |
Settling Time | Less than 5 seconds for all functions and ranges except as noted |
Standard Interfaces | IEEE-488 (GPIB), RS-232, 5725A (5500A only) |
Temperature Performance | Operating: 0°C to 50°C
Calibration: (tcal): 15°C to 35°C
Storage: -20°C to 70°C
|
Temperature Coefficient | Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification
(or 1-year, as applicable) per °C
|
Relative Humidity | Operating:
< 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage:
< 95%, non-condensing
|
Altitude | Operating:
3,050m (10,000 ft) maximum
Non-operating:
12,200m (40,000 ft) maximum
|
Safety | Designed to comply with:
IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92
|
Analog Low Isolation | 20V |
EMC | Designed to comply with FCC Rules Part 15 |
Line Power | Line Voltage (selectable):
100V, 120V, 220V, 240V
Line Frequency:
47 to 63 Hz
Line Voltage Variation:
± 10% about line voltage setting
|
Power Consumption | 300 VA |
Calibration Documentation | 17025 accredited report of calibration included |
Dimensions | Height:
17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width:
43.2 cm (17 in), standard rack width
Depth:
47.3 cm (18.6 in) overall
|
Weight (without options) | 20 kg (44 lbs.) |
Absolute Uncertainty Definition | 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration.
You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
|
Specification Confidence Interval | > 99% |
Specifications |
|
5500A Multi-Product Calibrator |
|
Model Overview |
|
Model | Description |
5500A | Multi-Product Calibrator |
Accessories |
|
55XX/CASE | Transit case with wheels |
5500A/COIL | 50-Turn Current Coil |
5500A/HNDL | Side handle |
5500A/LEADS | Test Lead Set |
5500A-SC300 | 300 MHz Oscilloscope Calibration Option |
5500A-SC600 | 600 MHz Oscilloscope Calibration Option |
5725A | Amplifier |
Y5537 | Rack mount kit |